Mineral insolated thermocouple assembly according or similar to DIN 43772
MIT with protection tube form 2

  • Note: also available as type-tested Ex i version in accordance with
    ATEX and IECEx for measuring in zone 0, 

    Type in Ex operating instructions manual TR15-B, Ex ia IIC T1...T6 Ga/Gb
    or Ex ia IIIC T135 °C Da/Db.
  • Process connection with separate stop flange or gas-tight threaded sleeve.
  • Area of application: Autoclaves, machine and plant construction, chemical, food and beverage industry, energy and power plant technology, sanitary, heating and air-conditioning technology.
  • Measuring range depending on thermocouple -50 to 1100°C.
  • Standard material of protective fitting 1.4571, preferred diameter 9 or 11 mm.
  • Measuring insert exchangeable in accordance with DIN 43735.
  • Thermocouples according to  IEC / EN 60584-1, insulated or welded to the floor.
  • Also with optional directly mounted transmitter.

Online request / Configurator

Plug in MIT model 2 with protection shell IP54 on request up to IP67

Your choice

Your note